Clean workstation testing

Camfil's NATA accredited facility can perform various compliance tests to validate clean workstations and laminar air flow cabinets, as per the following Australian Standards:

  • AS 1807.4.1 – Determination of air velocity and uniformity in clean work stations, laminar flow safety cabinets and pharmaceutical isolators
  • AS 1807.4.3 – Determination of work zone integrity
  • AS 1807.4.4 – Determination of integrity for terminally mounted HEPA Filter installations
  • AS/ISO 14644.1 – Particle Counting in work zone
  • AS 1807.4.5 – Determination of illuminance
  • AS 1807.4.7 – Determination of sound level at installed work stations, safety cabinets and pharmaceutical isolators
  • AS 1807.4.10 – Determination of intensity of radiation from UV lamps

At the conclusion of our comprehensive testing, you will receive a detailed analysis report and a NATA endorsed certificate of compliance to verify that your clean workstation or laminar air flow cabinet meets Australian Standard specifications and relevant state or industry requirements.

LAMINAR FLOW HOOD TESTING

Clean workstations and laminar flow hoods provide a controlled work surface through HEPA filtered air that is exhausted across a work surface in a laminar or unidirectional air stream creating a clean work zone.

Controlled workstations and laminar flow cabinets are often used within medical, pharmaceutical, scientific and manufacturing applications, including:

  • Aseptic dispensing
  • Cell culturing
  • Media pouring
  • Assembly of electronics and optical components
  • Preparation of medical devices

WHEN SHOULD CLEAN WORKSTATIONS BE TESTED?

  • On site, upon installation and/or commissioning, prior to use
  • After any mechanical or electrical maintenance
  • After filter replacement
  • After re-location of clean work stations or safety cabinets
  • Re-tested at least annually
  • Re-tested as required by appropriate regulatory authority
  • In special circumstances, such as a significant change in the work program, or where impaired operation is suspected

OUR ACCREDITATION INFORMATION

Our NATA accredited testing and certification services are provided under the following accreditation and site numbers:

  • Camfil Australia Pty Ltd - accreditation number 15330
  • Camfil Technical Services - Queensland - site number 16382
  • Camfil Technical Services - South Australia - site number 24043
  • Camfil Technical Services - Victoria - site number 24801
  • Camfil Technical Services - New South Wales - site number 24198
NATA Logo