Fume cupboard testing

Camfil's NATA accredited facility perform various compliance tests to validate fume cupboards and cabinets, as per the following Australian Standards:

  • AS 2243.8 – Safety in laboratories: Fume Cupboards (Appendices A and B)
  • AS 2243.9 – Safety in laboratories: Recirculating fume cabinets (Appendices A and B)

NATA accredited testing for ducted fume cupboards include:

  • AS/NZS 2243.8 App A – Smoke Test (Clause A6)
  • AS/NZS 22433.8 App B – Determining Face Velocity (Clause 3.1.2)
  • AS 1807.15 – Lighting Intensity (Clause 2.2.5)
  • AS 1807.20 – Sound Levels (Clause 4.2.4)

NATA accredited testing for recirculating fume cabinets include:

  • AS/NZS 2243.9 App A – Smoke Test (Clause A5)
  • AS/NZS 22433.9 App B – Determining Face Velocity (Clause 4.1.2)
  • AS 1807.15 – Lighting Intensity (Clause 2.3.6)
  • AS 1807.20 – Sound Levels (Clause 5.2.5)

At the conclusion of our comprehensive testing, you will receive a detailed analysis report and a NATA endorsed certificate of compliance to verify that your fume cupboard meets Australian Standard specifications.

When should fume cupboards be tested?

  • On site, prior to use, upon installation and/or commissioning
  • After any mechanical or electrical maintenance
  • Re-tested at six monthly
  • Re-tested as required by appropriate regulatory authority
  • In special circumstances, such as a significant change in the work program, or where impaired operation is suspected
Male scientist in a laboratory.

Our accreditation information

Our NATA accredited testing and certification services are provided under the following accreditation and site numbers:

  • Camfil Australia Pty Ltd - accreditation number 15330
  • Camfil Technical Services - Queensland - site number 16382
  • Camfil Technical Services - South Australia - site number 24043
  • Camfil Technical Services - Victoria - site number 24801
  • Camfil Technical Services - New South Wales - site number 24198
NATA logo