Megalam ProSafe

HEPA (H14) and ULPA (U15, U16, U17) filter panels especially made for turbulent and laminar airflow applications in high-tech cleanrooms, clean benches and clean air devices. Optimized for LifeScience and Food&Beverage industry with glass fiber media for low pressure drop and long life time

  • Prosafe certified for Pharma, Food & Beverage
  • 100% filter scan tested according to ISO29463
  • Chemically resistant to decontamination and cleaning agent
  • Microbial inert components acc. to ISO 846
  • Tested for food safety acc. to EC 1935:2004
  • Hygienic product acc. to VDI 6022 and ISO846
  • Free of bisphenol-A, phthalate and formaldehyde
  • Manufactured and packed in a controlled environment
  • ICON9 en1822
  • ICON20 prosafe
  • ICON42 UL logo
Item Type Filter Class EN1822 Width (mm) Height (mm) Depth (mm) Airflow (m3/h) Pressure drop Initial (Pa)
MD13- 305x610-10/22 H13 305 610 66 300 116
MD13- 610x610-10/22 H13 610 610 66 605 115
MD13- 914x610-10/22 H13 914 610 66 905 114
MD13- 1219x610-10/22 H13 1219 610 66 1205 114
MD13- 610x610x76-11/22 H13 610 610 76 605 118
MD13LS-305x610x71-10/22 H13 305 610 71 300 116
MD13LS-610x610x71-10/22 H13 610 610 71 605 115
MD13LS-914x610x71-10/22 H13 914 610 71 905 114
MD13LS-1219x610x71-10/22 H13 1219 610 71 1205 114
MD14-305x610-10/22 H14 305 610 66 300 150
MD14-610x610-10/22 H14 610 610 66 600 140
MD14-914x610-10/22 H14 914 610 66 905 140
MD14-1219x610-10/22 H14 1219 610 66 1205 140
MX14-305x610-10/22 H14 305 610 90 300 105
MX14-610x610-10/22 H14 610 610 90 605 95
MX14-914x610-10/22 H14 914 610 90 905 95
MX14-1219x610-10/22 H14 1219 610 90 1205 95
MG14-305x610-10/22 H14 305 610 110 300 75
MG14-610x610-10/22 H14 610 610 110 605 65
MG14-914x610-10/22 H14 914 610 110 905 64
MG14-1219x610-10/22 H14 1219 610 110 1205 64
MD15-305x610-10/22 U15 305 610 66 300 160
MD15-610x610-10/22 U15 610 610 66 605 145
MD15-914x610-10/22 U15 914 610 66 905 145
MD15-1219x610-10/22 U15 1219 610 66 1205 145
MX15-305x610-10/22 U15 305 610 90 300 125
MX15-610x610-10/22 U15 610 610 90 605 115
MX15-914x610-10/22 U15 914 610 90 905 115
MX15-1219x610-10/22 U15 1219 610 90 1205 115
MG15-305x610-10/22 U15 305 610 110 300 90
MG15-610x610-10/22 U15 610 610 110 605 80
MG15-914x610-10/22 U15 914 610 110 905 80
MG15-1219x610-10/22 U15 1219 610 110 1205 80

Specifications

Application
Hepa/ULPA filter for clean rooms and LAF benches
Filter Frame
Anodized aluminium
Media
Glass fiber
Max temperature (°C)
70 °C
Relative Humidity
100%
Recommended final pressure drop
2x Initial pressure drop
Separator
Hot-melt
Gasket
Silicone Gel, Seamless PU foamed gasket
Grille, Downstream
Expanded metal painted (RAL 9010)
Grille, Upstream
Expanded metal painted (RAL 9010)
Sealant
Polyurethane
Max Airflow
Nominal flow rate (if not, efficiency drops)
Max. final pressure drop
MD: 500 Pa, MX: 600 Pa, MG:800 Pa
Comment
All filter tested acc. to EN 1822 with individual protocol and packed in PE-film. Compliant with Prosafe requirements. Other editions on request.
Fire rating: UL 900

Related Products

Articles related to Megalam

HEPA FILTRATION-The Changing trends

HEPA filtration is required in a range of different Industries from Pharmaceuticals, Hospitals, Microelectronics and even the Food and Beverage Industry. Learn about the changing trends in HEPA filtration requirements, How HEPA filtration has advanced over the last number of years and how to identify and reduce the associated costs to your clean air requirements

Life science and healthcare Air quality

Cancer mortality risks in the environment exposed to air pollution & one formidable weapon in battle - air filtration

It’s no surprise that air pollution has also been associated with lung cancer. What is shocking, however, is that bad air can also increase the risk of mortality from other types of sarcoma, including breast, liver, and pancreatic cancer.

Air cleaners and air purifiers Air quality

Filter Replacement Strategies

Gas Turbine Application: The quality of air inlet filters is essential for optimal gas turbine performance. Protecting the turbine with site-specific filtration solutions will assure profits are kept high.

Energy and Power Systems

Call Camfil Middle East & Africa Regional Office!
Find a contact